Total Carbon
The sample is initially weighed into a tin “boat” which is folded and placed into the 29 position sample carousel. Upon operation, the tin boat containing the sample is dropped from the sample carousel into the sample valve. The sample valve is purged and the sample boat is dropped into the combustion tube where it is combusted in a high-temperature oxygen rich atmosphere. In that environment, all carbon within the sample is rapidly oxidized to CO2. Interfering reaction products (including sulfur oxides, halides, water and nitrous oxides) are removed by the post-combustion scrubbers. The resulting carbon dioxide is then swept into the CM5017 CO2 Coulometer where it is automatically measured using absolute coulometric titration.
Data Handling
Names, weights and sizes of up to 50 samples can be entered, to be used by the CM5017 in calculating the final result. Analytical progress is displayed on the 10” LCD touch screen in user-selectable units. Detailed analysis information is automatically saved to an on-board SD card after each sample. Data can also be transmitted through the standard serial and Ethernet ports to be captured on a personal computer or LIMS. In addition, a detailed report can be printed to the optional small format printer while each sample is running.
Ordering Information
CM220 – Total Carbon Autoanalyzer
Includes: CM5017 CO2 Coulometer and CM5200 Autosampler Furnace. (P/N CM220-01 110V, 50/60Hz) (P/N CM220-02 220V, 50/60Hz)
Optional Equipment:
Printer – 3” format impact printer. Includes cable, power supply, paper and ribbon. (P/N CM124078)