Available ViscometerTubes
· Cannon-Fenske Opaque
· Cannon-Fenske Routine
· Cannon-Ubbelohde Dilution
·
Cannon-Ubbelohde
Semi-Micro
·
Ubbelohde
· Zeitfuchs-Crossarm
Cannon-Fenske
Opaque (reverse flow type) viscometers are for liquids so dark in color that
the meniscus cannot be seen in Cannon-Fenske Routine viscometers. Cannon-Fenske
Opaque viscometers are used to study liquid asphalts and road oils at 60°C
(140°F) in the range from 30 to 6000 centistokes (see ASTM D-2170) and for studying the effects of various additives to
lubricating oils and hydraulic fluids at low temperature. External pressure can
be applied to these viscometers to vary shear stress and shear rate.
Cannon-Fenske
Opaque viscometers require a minimum sample volume of 12 mL and a liquid bath
depth of 229 mm (9 inches).
All
Cannon-Fenske Opaque viscometers conform to ASTM 446 and ISO 3105
specifications.
- Ordering Information –
|
|
|
Range
|
Viscometer |
Viscometer |
|
|
Size
|
(cSt) |
(Assembly) |
(Glassware) |
|
|
|
|
|
|
|
|
25 |
0.5 - 2 |
JI229-700-20 |
JI449-700-20 |
|
|
50 |
0.8 - 4 |
JI229-700-21 |
JI449-700-21 |
|
|
75 |
1.6 - 8 |
JI229-700-22 |
JI449-700-22
|
|
|
100 |
3-15 |
JI229-700-23 |
JI449-700-23 |
|
|
150 |
7-35 |
JI229-700-24 |
JI449-700-24 |
|
|
200 |
20-100 |
JI229-700-25 |
JI449-700-25 |
|
|
300 |
50-250 |
JI229-700-26 |
JI449-700-26 |
|
|
350 |
100-500 |
JI229-700-27 |
JI449-700-27 |
|
|
400 |
240-1200 |
JI229-700-28 |
JI449-700-28 |
|
|
450 |
500-2500 |
JI229-700-29 |
JI449-700-29 |
|
|
500 |
1600-8000 |
JI229-700-30 |
JI449-700-30 |
|
|
600 |
4000-20000 |
JI229-700-31 |
JI449-700-31 |
|
|
650 |
9000-45000 |
JI229-700-32 |
JI449-700-32 |
|
|
700 |
20000-100000 |
JI229-700-33 |
JI449-700-33 |
Cannon-Fenske
Routine viscometers are for measuring the kinematic viscosity of transparent
Newtonian liquids. Cannon-Fenske Routine viscometers are particularly well
suited for the study of light colored petroleum products and lubricants in
accordance with ASTM D-445 and ISO 3104. For darker liquids the use of
Cannon-Fenske Opaque viscometers is recommended.
Cannon-Fenske
Routine viscometers require a minimum sample volume of 7 mL and a liquid bath
depth of 203 mm (9 inches).
All
Cannon-Fenske Routine viscometers conform to ASTM 446 and ISO 3105
specifications.
- Ordering Information –
|
|
|
Range |
Viscometer |
Viscometer |
|
|
Size |
(cSt) |
(Assembly) |
(Glassware) |
|
|
|
|
|
|
|
|
25 |
0.5 - 2 |
JI229-710-01 |
JI449-710-01 |
|
|
50 |
0.8 - 4 |
JI229-710-02 |
JI449-710-02 |
|
|
75 |
1.6 - 8 |
JI229-710-03 |
JI449-710-03 |
|
|
100 |
3-15 |
JI229-710-04 |
JI449-710-04 |
|
|
150 |
7-35 |
JI229-710-05 |
JI449-710-05 |
|
|
200 |
20-100 |
JI229-710-06 |
JI449-710-06 |
|
|
300 |
50-250 |
JI229-710-07 |
JI449-710-07 |
|
|
350 |
100-500 |
JI229-710-08 |
JI449-710-08 |
|
|
400 |
240-1200 |
JI229-710-09 |
JI449-710-09 |
|
|
450 |
500-2500 |
JI229-710-10 |
JI449-710-10 |
|
|
500 |
1600-8000 |
JI229-710-11 |
JI449-710-11 |
|
|
600 |
4000-20000 |
JI229-710-12 |
JI449-710-12 |
|
|
650 |
9000-45000 |
JI229-710-13 |
JI449-710-13 |
|
|
700 |
20000-100000 |
JI229-710-14 |
JI449-710-14 |
Cannon-Ubbelohde
viscometers are a rugged version of the widely used Ubbelohde "suspended
level" design in which the lower meniscus is "suspended"
immediately below the working capillary to ensure a uniform driving head of
liquid that is independent of sample volume. With a viscometer constant that is
independent of temperature and typically requires no kinetic energy corrections
over the range from 0.5 to 100,000 centistokes, the improved Cannon-Ubbelohde
design is particularly well suited for use at temperatures above 93°C (200°F)
or below -18°C (0°F).
Cannon-Ubbelohde
Dilution viscometers require a minimum sample volume of 8.0 mL and a liquid
bath depth of 280 mm (11 inches). Initial samples can be diluted to as much as
40 mL for intrinsic viscosity determinations.
All
Cannon-Ubbelohde Dilution viscometers conform to ASTM 446 and ISO 3105
specifications.
- Ordering Information –
|
|
|
Range |
Viscometer |
Viscometer |
|
|
Size |
(cSt) |
(Assembly) |
(Glassware) |
|
|
|
|
|
|
|
|
25 |
0.5 - 2 |
JI229-300-01 |
JI449-300-01 |
|
|
50 |
0.8 - 4 |
JI229-300-02 |
JI449-300-02 |
|
|
75 |
1.6 - 8 |
JI229-300-03 |
JI449-300-03
|
|
|
100 |
3-15 |
JI229-300-04 |
JI449-300-04 |
|
|
150 |
7-35 |
JI229-300-05 |
JI449-300-05 |
|
|
200 |
20-100 |
JI229-300-06 |
JI449-300-06 |
|
|
300 |
50-250 |
JI229-300-07 |
JI449-300-07 |
|
|
350 |
100-500 |
JI229-300-08 |
JI449-300-08 |
|
|
400 |
240-1200 |
JI229-300-09 |
JI449-300-09 |
|
|
450 |
500-2500 |
JI229-300-10 |
JI449-300-10 |
|
|
500 |
1600-8000 |
JI229-300-11 |
JI449-300-11 |
|
|
600 |
4000-20000 |
JI229-300-12 |
JI449-300-12 |
Cannon-Ubbelohde
viscometers are a rugged version of the widely used Ubbelohde "suspended
level" design in which the lower meniscus is "suspended"
immediately below the working capillary to ensure a uniform driving head of
liquid that is independent of sample volume. With a viscometer constant that is
independent of temperature and typically requires no kinetic energy corrections
over the range from 0.5 to 100,000 centistokes, the improved Cannon-Ubbelohde
design is particularly well suited for use at temperatures above 93°C (200°F)
or below -18°C (0°F).
Cannon-Ubbelohde
Semi-Micro viscometers require a minimum sample volume of 1.0 mL (sizes 400 -
600 may require more) and a liquid bath depth of 240 mm (9.5 inches). Initial
samples can be diluted to as much as 20 mL for intrinsic viscosity
determinations.
All
Cannon-Ubbelohde Semi-Micro viscometers conform to ASTM 446 and ISO 3105
specifications.
- Ordering Information –
|
|
|
Range |
Viscometer |
Viscometer |
|
|
Size |
(cSt) |
(Assembly) |
(Glassware) |
|
|
|
|
|
|
|
|
25 |
0.5 - 2 |
JI229-100-01 |
JI449-100-01 |
|
|
50 |
0.8 - 4 |
JI229-100-02 |
JI449-100-02 |
|
|
75 |
1.6 - 8 |
JI229-100-03 |
JI449-100-03 |
|
|
100 |
3-15 |
JI229-100-04 |
JI449-100-04 |
|
|
150 |
7-35 |
JI229-100-05 |
JI449-100-05 |
|
|
200 |
20-100 |
JI229-100-06 |
JI449-100-06 |
|
|
300 |
50-250 |
JI229-100-07 |
JI449-100-07 |
|
|
350 |
100-500 |
JI229-100-08 |
JI449-100-08 |
|
|
400 |
240-1200 |
JI229-100-09 |
JI449-100-09 |
|
|
450 |
500-2500 |
JI229-100-10 |
JI449-100-10 |
|
|
500 |
1600-8000 |
JI229-100-11 |
JI449-100-11 |
|
|
600 |
4000-20000 |
JI229-100-12 |
JI449-100-12 |
Ubbelohde
viscometers are a less rugged version of the Cannon-Ubbelohde "suspended
level" design in which the lower meniscus is "suspended"
immediately below the working capillary to ensure a uniform driving head of
liquid that is independent of sample volume. With a viscometer constant that is
independent of temperature and typically requires no kinetic energy corrections
over the range from 0.5 to 100,000 centistokes, the Ubbelohde design is
particularly well suited for use at temperatures above 93°C (200°F) of below
-18°C (0°F).
Ubbelohde
viscometers require a minimum sample volume of 11 mL and a liquid bath depth of
241 mm (9.5 inches).
All
Ubbelohde viscometers conform to ASTM 446 and ISO 3105 specifications.
- Ordering Information –
|
|
|
Range |
Viscometer |
Viscometer |
|
|
Size |
(cSt) |
(Assembly) |
(Glassware) |
|
|
|
|
|
|
|
|
0 |
0.3 - 1 |